GY481 portable three-phase power quality analyzer
GYZH-II CT Volt-Amp Ratio Polarity Comprehensive Te
SMG2000B digital double clamp phase voltammeter
SMG3000 three-phase phase voltammeter
GYRLC-8YQ Gas Relay Calibrator
GY2012B Gas Relay Calibration Station
DNC three-phase electricity tester
GYZH-I CT Volt-Amp Ratio Polarity Comprehensive Tes
GY9500 high voltage CT ratio tester (online)
Technical Specifications:
Ambient temperature |
(23±5)℃ |
|
environment humidity |
(45~75)% RH |
|
Signal waveform under test |
Sine wave, β = 0.02 |
|
Test signal frequency |
(50±0.2)Hz |
|
The position of the current-carrying conductor under test in the jaws |
no limit |
|
Amplitude range of measured signal when measuring phase |
100~220V、0.5~1.5A |
|
External reference frequency electromagnetic field interference |
Should be avoided |
|
AC voltage measurement error |
|
|
Limit |
Resolution |
Tolerance |
20V |
0.01V |
±(1.2%RD+2) |
200V |
0.1V |
±(1.0%RD+2) |
500V |
1V |
±(1.2%RD+2) |
Rated operation conditions |
|
|
Ambient temperature |
(0~40)℃ |
|
environment humidity |
(20~80)% RH |
|
Signal waveform under test |
Sine wave, β = 0.05 |
|
Test signal frequency |
(50±0.5)Hz |
|
The position of the current-carrying conductor under test in the jaws |
no limit |
|
Amplitude range of measured signal when measuring phase |
phase test U1-U2:30V~500V |
|
phase test I1-I2 :10mA~10.00A |
|
|
phase test U1-I2 or I1-U2 :10V~500V、10mA~10.00A |
|
|
External reference frequency electromagnetic field interference |
Should be avoided |
|
Rated working error limit |
Under the above-mentioned rated working conditions, the measured rated working error limit does not exceed twice the corresponding basic error limit |
|
Power frequency phase measurement error (voltage input loop impedance when measuring U1-U2 phase: 40KΩ) |
|
|
range |
Resolution |
Tolerance |
0~360° |
1° |
±3° |
|
|
|
|
|
|
Other technical characteristics |
|
|
Display digits |
Three and a half |
|
Sampling rate |
3 times / sec |
|
power supply |
Single 9V stacked battery, power supply current is less than 5mA |
|
Dimension |
Case size |
192×95×55mm3 |
Clamp shell size |
140×42×20mm3 |
|
Jaw size |
Φ7mm×9mm |
|
weight |
Body |
280g |
|
Measuring pliers |
2×200g |
Storage conditions |
temperature |
-10℃~50℃ |
Products Features:
1. Withstand voltage: between the voltage input terminal and the case, between the iron core of the
clamp-shaped current transformer (current clamp) and the clamp handle and the secondary winding
coil, it can withstand 1000V / 50Hz, and the two voltage input terminals can withstand 500V / 50Hz
sine wave AC voltage test for 1min