GY3201 Automatic Closed Flash Point Tester

Recommended Products
Product Details
      Technical Specifications:

          Range              Room temperature ~ 300 ℃
     Repeatability                                 0.50%
      Resolution                                 0.1 ℃
      Precision                                 0.50%
Temperature sensor              Platinum resistance (PT100)
     Flash sensor                       Ion detection ring
Ambient temperature                             10 ~ 40 ℃
  Relative humidity                                 <85%
    Supply voltage                         AC220V ± 10%
          Power                               <500W
      Heating rate Comply with national standards, the United States, the European Union and other standards
      Dimensions               Host: 190 × 260 × 285 (mm)
      Heating furnace: 280 × 260 × 285 (mm)
  Instrument weight                                 18kg

          Product Features:

1. Adopt a new type of high-speed digital signal processor, work reliably and with high precision
2. Detection, cover opening, ignition, alarm, cooling, printing, the whole experiment process is completed automatically
3. Platinum heating wire (standard), gas ignition method (optional)
4. Automatic detection of atmospheric pressure and automatic correction of test results
5. Using newly developed high-power high-frequency switching power supply heating technology, high heating efficiency,
    using adaptive PID control algorithm, automatically adjusting the heating curve
6. Automatic detection and alarm when the temperature exceeds the value
7. Thermal miniature printer makes printing more beautiful and faster, with offline printing function
8. Historical records with time stamps, up to 255
9. The 100-year calendar clock with temperature compensation, accurate travel time, automatically records the date and
     time of measurement, and can run for more than 10 years in the state of power failure
10. Use 320x240 large-screen graphic LCD display, Chinese character display interface, rich content
11. Using touch screen buttons, intuitive and convenient operation
12. Multiple built-in execution standards are available
 