WhatsApp:

87197008

Email Address:

contact@guoyi.com.sg

Tel:+65-93262640

HP:+65-87197008

Home > PRODUCTS > Grounding & Insulation INSTR. > Grounding Test INSTR >

BY2571 Digital Ground Resistance Tester

Recommended Products
Product Details
 

Technical Specifications

   
 

       Conditions of Use

Ambient temperature

0+45

 

Relative humidity

≤85%RH

 

 Measuring range and constant current value (RMS)

resistance

02Ω(10mA)220Ω(10mA)20200Ω(1mA)

 

Voltage

AC019.99V

 

 Measurement accuracy and resolution

Accuracy

00.2Ω≤±3%±1d

 

0.2Ω199.9Ω≤±1.5%±1d

 

119.99V≤±3%±1d

 

Resolution

0.001Ω0.01Ω0.1Ω0.01

 

 Measurement error caused by grounding resistance and ground voltage

Allow auxiliary ground resistance

RC (between C1 and C2)

 

0 2Ω, 2 20Ω≤1kΩ

 

20 200Ω≤2kΩ

 

RP (between P1 and P2) <40kΩ, error ≤ ± 5%

 

Allowable ground voltage (power frequency RMS)

≤5V, error ≤ ± 5%

 

 Power supply and power consumption

Maximum power loss

≤2W

 

DC

8 × 1.5V (AA, R6) battery

 

AC

220V / 50Hz

 

 Volume and weight

volume

220 × 200 × 105mm3

 

weight

≤1.4kg

      Product Features:

1. The structure of the grounding resistance measuring instrument adopts high-strength aluminum alloy as the chassis. The circuit uses a phase-locked
    loop synchronous tracking detection method to prevent power frequency and radio frequency interference and is equipped with a switched capacitor
   filter to make the instrument have better anti-interference ability

2. Adopt DC / AC conversion technology to change DC to AC low-frequency constant current for easy measurement

3. Allow the auxiliary grounding resistance to change between 0 1.8KΩ (RC) and 0 40KΩ (RP), so as not to affect the measurement result

4. The ground resistance measuring instrument does not need to manually adjust the balance. The 3 (1/2) digit LCD display, in addition to measuring
   the ground resistance, can also measure low resistance conductor resistance, soil resistivity and AC ground voltage. If the test circuit fails to display
   "1" in the header, it means overflow, which is in line with the conventional measurement habits