DNC three-phase energy meter field calibrator
GY481 Three-phase precision programmable power supp
GY-HG Intelligent Transformer Calibrator
GY9201 AC and DC indicator verification device
GY4501 single-phase energy meter field calibrator
Transformer load box
GYCT standard current transformer
GY series current riser
GYPT standard voltage transformer
GY4500 Secondary Voltage Drop Tester
Technical Specifications
Equipment use environment |
Temperature: 5 ℃ ~ 40 ℃ |
Relative humidity: <80% (at 25 ℃) |
|
Altitude: <2500m |
|
||
Power frequency: 50Hz ± 0.5Hz |
Power supply voltage: 220V ± 5V |
||
Measuring range |
In-phase component (%): 0.0001 ~ 200.0 |
Resolution: 0.0001 |
|
Orthogonal component (min): 0.001 ~ 700.0 |
Resolution: 0.001 |
||
Impedance (Ω): 0.0001 ~ 20.0 |
Resolution: 0.0001 |
||
Admittance (ms): 0.0001 ~ 20.0 |
Resolution: 0.0001 |
||
Basic error |
In-phase component |
ΔX = ± (X × 2% + Y × 2%) ± Dx (with 1 level optional) |
|
Orthogonal component |
ΔY = ± (X × 2% + Y × 2%) ± Dy (with 1 level optional) |
||
"X", "Y"-display value of the instrument |
|||
"Dx, Dy" —— Instrument quantization error |
|||
Dx = 2, Dy = 5 |
|||
Dial indicator |
class 2(class 1 available) |
||
Operating range |
Current |
(1%~149%)In (In =5A) |
|
(5%~149%)In (In =1A) |
|||
Voltage |
(5%~149%)Un (Un =100V,150V,100V/) |
||
(5%~149%)Un (Un =100V/3) |
|||
Workload |
Electric current |
TO to TX<0.12Ω cosΦ=1 |
|
Voltage |
a to x<0.25VA (100V) |
||
Polarity error indication |
When the operating current (voltage) reaches more than 5% of the rated current (voltage) and the error exceeds 180%, there should be a polarity indication. |
||
Note: If it is more than 10% of the rated operating current (voltage), the polarity indication should not appear, indicating that there is a fault, please do not increase the current (voltage), so as not to burn the instrument. |
|||
Ratio error indication |
When the operating current (voltage) reaches more than 5% of the rated current (voltage) and the error exceeds 30% but less than 180%, there should be a ratio error indication. |
||
Insulation and voltage test and description |
Ground Terminal and TX () terminal is connected |
||
The power socket can bear 1.5kV to the shell, withstand voltage for 1m |
Product Features
1. The intelligent transformer calibrator uses 320 × 240 dot-matrix liquid crystal, with a large viewing range and long-life
backlight illumination, which is convenient to use
2. Full Chinese operation interface, beautiful interface
3. The dial indicator, ratio difference and angle difference are displayed in extra large font, which is convenient for users to observe
4. The intelligent transformer calibrator produced by our company fully meets the requirements of JJG314-1993 and JJG314-1994,
and automatic sampling
5. Automatic judgment of wrong polarity and wrong ratio
6. The host computer software has the function of opening virtual instruments
7. Completely solve the S-level measurement problem
8. Automatic switching range
9. Adopt the perfect combination of advanced unique circuit and DSP technology to completely remove the problem of instability
of the resistance-capacitance phase-shift circuit
10. Power consumption: <15VA (without micro printer); <25VA (with micro printer)
11. Harmonic suppression ratio:> 40db
12. Overall dimensions (L × W × H): 260 × 350 × 150mm3
13. Weight: 6kg